Opportunity
SAM #SPMYM126Q4012
One HLI 9000 MK IV dew point indicator for Norfolk Naval Shipyard
Buyer
DLA Maritime - Norfolk
Posted
November 19, 2026
Respond By
November 25, 2026
Identifier
SPMYM126Q4012
NAICS
333248
DLA Maritime Norfolk is seeking a new dew point indicator for Norfolk Naval Shipyard shipboard use. - Government buyer: Defense Logistics Agency (DLA), DLA Maritime Norfolk, for Norfolk Naval Shipyard (NNSY); buyer contact: Ursula Cruz. - Requested product: One HLI 9000 MK IV dew point indicator, model HLI 900CG504, part number 900CG504. The notice does not identify the manufacturer or a vendor by company name. - Measurement requirements: Dew-point range of -60°C to +20°C and accuracy of ±2°C; display resolution of 0.1 and measurement resolution of 0.01°C. Repeatability must be better than 0.1°C, stability 0.1°C, and hysteresis 0.05°C. Supported units include °F, °C, and K for dew point and gas temperature, plus ppmv, ppmw, %RH, g/m³, and g/kg for moisture measurements. - Operating and construction requirements: Outdoor operation at 0–100% relative humidity, including condensing conditions, and -20°C to +50°C; storage and transport from -40°C to +70°C. The internal sensor must support up to 5,000 psig; higher pressures require external sampling. Required features include a rechargeable internal battery with 115 V AC charging, blue LCD, aluminum enclosure, shipboard-compatible sample-port connections, a 40-micron stainless-steel inlet filter, and AISI 316L stainless-steel gas-wetted materials. - Other requirements: Sensor flow of 1–9.4 nlpm (2–20 scfb), stated compact dimensions and weight under 6.6 lb (3 kg), and an English user manual. The unit must be new; used or refurbished equipment is not accepted.
Description
PLEASE FILL OUT ATTACHMENT AND RETURN WITH OFFER
• HLI 9000 MK IV
With the following configuration:
■ Measurement accuracy: ±2°C from -60° to +20° dew point.
■ Measurement Range: -60° to +20° dew point.
■ Measurement units: °F, °C, °K for dew point and gas temperature. ppmv, ppmw for air, N2, H2, CO2, SF6, %RH, g/m3, g/kg.
■ Resolution (display): 0.1 for all dew-point derived units and auto ranging where appropriate ( e.g. ppm).
■ Resolution (measurement): 0.01 ° C (0.02° F) dew point.
■ Measurement system repeatability: Better than 0.1 ° C (0.18° F).
■ Measurement system stability: 0.1° C (0.18° F).
■ Measurement system hysteresis: 0.05° C (0.09° F).
■ Electrical input/output:
• Power supply: 115 V AC power to recharge the internal battery.
■ Operating conditions:
• Operating Environment: Outdoors, 0 to + 100% RH condensing; -20° to
+ 50°C (-4° to+ 122° F).
• Storage/fransport Temperature: -40° to +70° C (-40° to+ 158° F).
• Operating Pressure Range: 5000 psig for internal sensor, higher pressures with external sampling ■ Mechanical specifications:
• Display: Blue LCD graphical display.
• Enclosure type: Aluminum.
• Gas connections: Compatible with existing shipboard installed sample ports; other options available.
• Flow across sensor: 1 to 9.4 nlpm (2 to 20 scfb).
• Filter: 40 micron stainless steel sinter in the inlet port.
• Gas wetted materials: AISI 316L stainless steel.
• Outline dimensions: 17 3/16 in x 12 5/32 in x 6 7/32 in (437 mm x 309 mm x 158 mm).
• Weight: Hygrometer: <6.6 lb (3 kg).
■ User Manual English systems.