# SAM #SPMYM126Q4012 

One HLI 9000 MK IV dew point indicator for Norfolk Naval Shipyard

**Buyer:** DLA Maritime - Norfolk
**Posted:** November 19, 2026
**Respond By:** November 25, 2026
**Identifier:** SPMYM126Q4012
**NAICS:** 333248

DLA Maritime Norfolk is seeking a new dew point indicator for Norfolk Naval Shipyard shipboard use.
- **Government buyer:** Defense Logistics Agency (DLA), DLA Maritime Norfolk, for Norfolk Naval Shipyard (NNSY); buyer contact: Ursula Cruz.
- **Requested product:** One HLI 9000 MK IV dew point indicator, model HLI 900CG504, part number 900CG504. The notice does not identify the manufacturer or a vendor by company name.
- **Measurement requirements:** Dew-point range of -60°C to +20°C and accuracy of ±2°C; display resolution of 0.1 and measurement resolution of 0.01°C. Repeatability must be better than 0.1°C, stability 0.1°C, and hysteresis 0.05°C. Supported units include °F, °C, and K for dew point and gas temperature, plus ppmv, ppmw, %RH, g/m³, and g/kg for moisture measurements.
- **Operating and construction requirements:** Outdoor operation at 0–100% relative humidity, including condensing conditions, and -20°C to +50°C; storage and transport from -40°C to +70°C. The internal sensor must support up to 5,000 psig; higher pressures require external sampling. Required features include a rechargeable internal battery with 115 V AC charging, blue LCD, aluminum enclosure, shipboard-compatible sample-port connections, a 40-micron stainless-steel inlet filter, and AISI 316L stainless-steel gas-wetted materials.
- **Other requirements:** Sensor flow of 1–9.4 nlpm (2–20 scfb), stated compact dimensions and weight under 6.6 lb (3 kg), and an English user manual. The unit must be new; used or refurbished equipment is not accepted.

### Description

<p>PLEASE FILL OUT ATTACHMENT AND RETURN WITH OFFER</p>

<p>&bull;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp; HLI 9000 MK IV</p>

<p>With the following configuration:</p>

<p>■&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp; Measurement accuracy: &plusmn;2&deg;C from -60&deg; to +20&deg; dew point.</p>

<p>■&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp; Measurement Range: -60&deg; to +20&deg; dew point.</p>

<p>■&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp; Measurement units: &deg;F, &deg;C, &deg;K for dew point and gas temperature. ppmv, ppmw for air, N2, H2, CO2, SF6, %RH, g/m3, g/kg.</p>

<p>■&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp; Resolution (display): 0.1 for all dew-point derived units and auto ranging where appropriate ( e.g. ppm).</p>

<p>■&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp; Resolution (measurement): 0.01 &deg; C (0.02&deg; F) dew point.</p>

<p>■&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp; Measurement system repeatability: Better than 0.1 &deg; C (0.18&deg; F).</p>

<p>■&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp; Measurement system stability: 0.1&deg; C (0.18&deg; F).</p>

<p>■&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp; Measurement system hysteresis: 0.05&deg; C (0.09&deg; F).</p>

<p>■&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp; Electrical input/output:</p>

<p>&bull;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp; Power supply: 115 V AC power to recharge the internal battery.</p>

<p>■&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp; Operating conditions:</p>

<p>&bull;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp; Operating Environment: Outdoors, 0 to + 100% RH condensing; -20&deg; to</p>

<p>+&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp; 50&deg;C (-4&deg; to+ 122&deg; F).</p>

<p>&bull;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp; Storage/fransport Temperature: -40&deg; to +70&deg; C (-40&deg; to+ 158&deg; F).</p>

<p>&bull;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp; Operating Pressure Range: 5000 psig for internal sensor, higher pressures with external sampling ■ &nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp; Mechanical specifications:</p>

<p>&bull;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp; Display: Blue LCD graphical display.</p>

<p>&bull;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp; Enclosure type: Aluminum.</p>

<p>&bull;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp; Gas connections: Compatible with existing shipboard installed sample ports; other options available.</p>

<p>&bull;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp; Flow across sensor: 1 to 9.4 nlpm (2 to 20 scfb).</p>

<p>&bull;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp; Filter: 40 micron stainless steel sinter in the inlet port.</p>

<p>&bull;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp; Gas wetted materials: AISI 316L stainless steel.</p>

<p>&bull;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp; Outline dimensions: 17 3/16 in x 12 5/32 in x 6 7/32 in (437 mm x 309 mm x 158 mm).</p>

<p>&bull;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp; Weight: Hygrometer: &lt;6.6 lb (3 kg).</p>

<p>■ &nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp; User Manual English systems.</p>

[View original listing](https://sam.gov/opp/2e2556196d3b41ea8671cd1faaf9c737/view)