Opportunity
SAM #36C25227Q0107
Madison VA CT, MRI, magnet, and chiller full-service maintenance
Buyer
VA Contracting Office 252
Posted
November 23, 2026
Respond By
December 02, 2026
Identifier
36C25227Q0107
NAICS
811210
The Department of Veterans Affairs is seeking capability information for a proposed sole-source maintenance contract at the Madison VA. - Buyer: VA Great Lakes Acquisition Center (GLAC), Contracting Office 36C252. - Proposed contractor: GE Precision Healthcare LLC; the notice states the VA intends to negotiate a sole-source, firm-fixed-price contract. It is not a request for competitive quotes. - Equipment covered: One GE Discovery CT750-64 CT; one GE 1.5T MRI; one GE 3T MRI; two GE MR magnet maintenance and cryogen systems; and two Dimplex chillers (20-ton and 30-ton), all at William S. Middleton Memorial Veterans Hospital in Madison, Wisconsin. - Service scope: Full-service preventive maintenance and emergency repairs, including labor, travel, parts, materials, tools, software and licensing, updates, patches, and service documentation. - Key requirements: At least 95% operational uptime; preventive maintenance at least semiannually; phone response within two hours; and on-site response within two hours when phone troubleshooting is insufficient, as specified in the notice. Service must follow manufacturer requirements and applicable standards. - Response opportunity: Interested firms believing they can meet the requirements may submit capability information and customer references. - Term: 12-month base period plus four 12-month option periods.
Description
Under Statutory Authority 41 U.S.C. 4106(c), the VA Great Lakes Acquisition Center (GLAC) intends to negotiate a sole source contract with GE Precision Healthcare LLC for full-service maintenance and emergency repairs on the imaging equipment and software (see SOW below).
This acquisition is conducted under the authority of 41 U.S.C. 4106(c), order against multiple award schedule, as implemented in FAR 16.507-4, only one responsible source and no other supplies or services will satisfy agency requirements. This will be a firm fixed price contract for base plus four option periods. The NAICS Code is 811210.
This Notice of Intent is not a request for competitive quotes. No solicitation documents are available and telephone requests will not be honored. However, if a firm believes it can meet the requirements, it must furnish information about its products and services, as well as references from other customers who are using these products and services, to the Contracting Officer by 9:00 AM (local time) 12/02/2026. Supporting evidence must be furnished in sufficient detail to demonstrate the ability to comply with the above requirements. Responses received will be evaluated; however, a determination by the Government not to compete the proposed procurement based upon responses to this notice is solely within the discretion of the Government. If no responses are received, the Contracting Officer will proceed with the sole source negotiation with GE Precision Healthcare LLC.
Capability statements shall be submitted by email only as an MS Word or Adobe PDF attachment to Lori Eastmead at Lori.Eastmead@va.gov. TELEPHONIC INQUIRIES WILL NOT BE ACCEPTED.
STATEMENT OF WORK/SPECIFICATIONS
SCOPE: Contractor shall provide labor, travel expenses, parts, material, tools, software and licensing, and equipment necessary to provide full-service maintenance and emergency repairs of the G.E. equipment and software outlined in Attachment A below. Equipment is located at William S. Middleton, Madison VA, 2500 Overlook Terrace, Madison, WI 53705, VISN 12.
Attachment A Room: DG237 (Discovery) CT750-64EE#: 47959Site#-Sys ID: 608280 CT750Serial #: Gantry: 441638CN4 Room: DG500 1.5Tesla MRIEE#: 52415Site#-Sys ID: 608VAARTISTSerial #: HM17998 Room: DG513 GE MR MAGNET MAINTENANCE AND CRYOGEN (MSC28Z)Site#-Sys ID: 608VAARTISTC DIMPLEX MV PR DIMPLEX WO2-2-5000 CHILLER 20T (SDI020)EE#: 76668Site#-Sys ID: 608VAARTISTCHSerial #: W136510-001 Room: DG245B 3Tesla MRIEE#: 47960Site#-Sys ID: 608VAMR750WSerial #: UA0436 Room: DG245C GE MR MAGNET MAINTENANCE AND CRYOGEN (MSC28Z)Site#-Sys ID: 608VAMR750WC DIMPLEX MV PR DIMPLEX WO2-2-7500 CHILLER 30T (SDI030)EE#: 76667Site#-Sys ID: 608VAMR750WCH1Serial #: W006937-001
NOTE: Software includes all components of software that are available for coverage. Coverage shall include all technical aspects and all software and enterprise, and base aspects related to the product. Hardware shall be excluded for coverage relating to software.
CONTRACT PERIOD: Contract shall be a base period (12 months) with renewal options for four (4) additional periods to be exercised at the discretion of the government.
CONFORMANCE STANDARDS: Contract service shall ensure that the equipment functions in conformance with the latest published edition of applicable industry standards, including but not limited to the National Electrical Code (NEC), National Fire Protection Association (NFPA-99), Underwriters Laboratories (UL), Occupational Safety and Health Administration (OSHA), VA, Center for Devices and Radiological Health (CDRH), and federal, state, and local codes. Equipment defined herein shall be maintained in accordance with the manufacturer's standards/specifications. Preventive maintenance inspection/repair procedures shall be in accordance with the manufacturer's specifications and schedule for preventive maintenance (PM). The C&A requirements do not apply, and a Security Accreditation Package is not required.
OPERATIONAL UPTIME REQUIREMENTS: The equipment shall be operable and available for use 95% of the normal operating hours of the equipment as detailed in Paragraph 4, HOURS OF COVERAGE. Downtime shall be computed from notification of problems during normal work hours. Scheduled maintenance shall be excluded from downtime. Normal work hours are listed below, Monday through Friday, excluding national holidays. Operational uptime shall be computed during a month-long period. Repeated failure to meet this requirement can subject the contractor to DEFAULT action.
HOURS OF COVERAGE: Normal hours of coverage are Monday through Friday from 8:00 am to 6:00 pm (local time) for all facilities (excluding holidays). All service/repairs will be performed during normal hours of coverage unless requested or approved by the Contracting Officer (CO), Contracting Officer's Representative (COR), or their designee upon prior notification to the CO. If work must be performed outside of normal business hours, the contractor shall obtain prior approval from the COR or designee. If such work falls on a national holiday (Christmas, New Year's Day, Memorial Day, Independence Day, Labor Day, or Thanksgiving), service shall be paid separately from this contract. These hours shall be charged at the prevailing contractor non-business travel and labor rates. An obligation number shall be obtained prior to commencement of service. Service performed outside of the normal hours of operation at the request of the Contractor field service engineer (FSE) shall be considered service during normal hours of coverage. Work outside of prescribed hours shall be approved by the COR or designee.
Federal holidays observed by the VA are: New Year's Day Martin Luther King Day Presidents' Day Memorial Day Juneteenth Independence Day Labor Day Columbus Day Veterans' Day Thanksgiving Day Christmas Day Any other day specifically declared by the President of the United States to be a National or Federal Holiday.
B. Preventive maintenance shall be performed at least semi-annually, but more often if required to keep equipment operating to original manufacturer specifications. Preventive maintenance shall be scheduled with the COR at a mutually agreeable time during contract coverage hours.
NOTE: IF SERVICES ARE REQUIRED AFTER HOURS, THE CONTRACTOR SHALL BE REQUIRED TO PROVIDE A WRITTEN OFFER PRIOR TO ANY SERVICES BEING PERFORMED FOR APPROVAL BY THE COR. NOTE: Hardware/software update/upgrade installations shall be scheduled and performed outside normal hours of coverage at no additional charge to the Government (unless it would be detrimental to equipment up-time; to be determined by the COR).
UNSCHEDULED MAINTENANCE (Emergency Repair Service): Contractor shall maintain the equipment in accordance with the Conformance Standards section or the manufacturer's specifications. The Contractor shall provide repair service which may consist of calibration, cleaning, oiling, adjusting, replacing parts, and maintaining the equipment, including all intervening calls necessary between regular services and calibrations. All required parts shall be furnished. The COR or designee has the authority to approve/request a service call from the Contractor. Response Time: Contractor's FSE shall respond with a phone call to the COR or their designee within 2 hours after receipt of telephone notification. If the problem cannot be corrected by phone, the FSE shall commence work (on-site physical response) within 2 hours after receiving this second notification and shall proceed progressively to completion without undue delay.
SCHEDULED MAINTENANCE: The Contractor shall perform PM service to ensure that equipment listed in the schedule performs in accordance with the Conformance Standards and the manufacturer's specifications. The Contractor shall provide and utilize procedures and checklists with worksheet originals indicating work performed and actual values obtained (as applicable) provided to the COR at the completion of the PM. The Contractor shall provide a written description of Preventive Maintenance Inspections (PMI). This description shall include an itemized list of the procedures performed, including electrical safety. PM services include, without limitation, the following: Cleaning of equipment. Reviewing operating system software diagnostics to ensure that the system is operating to the manufacturer's specifications. Calibrating and lubricating the equipment. Performing remedial maintenance of a non-emergent nature. Testing and replacing faulty and worn parts and/or parts likely to become faulty, fail, or become worn. Inspecting and replacing, where indicated, electrical wiring and cables for wear and fraying. Inspecting and replacing, where indicated, all mechanical components, which may include, but are not limited to: X-ray tube mounting hardware, patient restraints and support devices, cables and mounting hardware, chains, belts, bearings and tracks, interlocks, clutches, motors, keyboards, and patient couches, for mechanical integrity, safety, and performance. Returning the equipment to the operating condition in accordance with the Conformance Standards and manufacturer's specifications. Providing documentation of services performed. Inspecting and calibrating the hard copy image device. PM services shall be performed in accordance with, and during the hours defined in, the preventive maintenance schedule established herein. All exceptions to the PM schedule shall be arranged and approved in advance with the COR. Any charges for parts, services, manuals, tools, or software required to successfully complete scheduled PM are included within this contract and its agreed-upon price, unless specifically stated in writing otherwise. Contractor shall provide all software updates and new releases to the systems. Contractor shall provide software patches and bug fixes to avoid or correct an error in the supported systems.
PARTS: The Contractor shall furnish and replace parts to meet uptime requirements. The Contractor has ready access to all parts, including unique and/or high-mortality replacement parts. All parts supplied shall be compatible with existing equipment. The contract shall include all parts. The Contractor shall use new or OEM-sanctioned rebuilt or refurbished parts. All parts shall be of current manufacture and have complete versatility with the presently installed equipment. All parts shall perform identically to the original equipment specifications. Rebuilt parts, used parts, or those removed from another piece of equipment shall not be installed without specific approval by the COR. Should the VA choose to provide its own parts that meet specifications, the Contractor shall install these parts as part of this contract.
SERVICE MANUALS/TOOLS/EQUIPMENT: The VA shall not provide tools, test equipment, service manuals, or service diagnostic software to the Contractor. The Contractor shall obtain, have on file, and make available to its FSEs all operational and technical documentation (such as operational and service manuals, schematics, and parts lists) necessary to meet the performance requirements of this contract. The location and listing of the service data manuals, by name and/or the manuals themselves, shall be provided to the COR upon request.
DOCUMENTATION/REPORTS: The documentation shall include detailed descriptions of the scheduled and unscheduled maintenance (i.e., emergency repairs) procedures performed, including replaced parts and prices (for outside normal working hours) required to maintain the equipment in accordance with conformance standards or the manufacturer's specifications. Such documentation shall meet the guidelines set forth in the conformance standards section. The Electronic Service Report (ESR) shall consist of a separate PM report for the item(s) covered under the “specific” contract. Grouping different equipment from different contracts on “one” ESR is prohibited. In addition, each ESR shall, at a minimum, document the following data legibly and in complete detail: Name of Contractor and contract number. Name of FSE who performed services. Contractor service ESR number/log number. Date, time (starting and ending), equipment downtime, and hours on-site for service call. VA purchase order number(s) covering the call if outside normal working hours. Description of problem reported by COR/user (if applicable). Identification of equipment to be serviced: inclusive of VA Asset VA (EE #), invoice number, manufacturer's name, device name, model number, serial number, and any other manufacturer's identification numbers. Itemized description of service performed (including, if applicable, costs associated with after-normal-working-hour services), including labor and travel, parts (with part numbers), materials, and circuit location of problem/corrective action. Total cost to be billed (if applicable—i.e., part(s) not covered, or service rendered after normal hours of coverage). Signatures: FSE performing services described. VA employee who witnessed service described. NOTE: ANY ADDITIONAL CHARGES CLAIMED SHALL BE APPROVED BY THE COR BEFORE SERVICE IS COMPLETED! FOOTNOTE: The Contractor may design/develop an ESR report form that incorporates the above minimum criteria. This report shall be approved by the COR prior to use and implementation. The report shall be submitted on the same day of service, or no later than one (1) business day, either electronically or on paper. If the ESR is solely digital, then the same minimum criteria are required.
FACILITY CHECK IN/CHECK OUT REQUIREMENTS: Vendor shall check in with the COR and/or designee prior to beginning any work. Due to the nature of working at a Government facility, the Contractor's staff shall be required to sign in and out each day with Biomedical Engineering. This sign-in is MANDATORY.
Contact: Healthcare Technology Management Location: Building 1 (Main Hospital), C143 (Ground Floor) Telephone Number:
Contractor FSEs shall always wear visible identification while on the premises of the Madison VA Hospital and follow all local visitor badging policies. Outside of normal business hours, Contractor shall check in at the police desk to obtain a contractor badge before entering the work site. The Contractor shall be required to report to Biomedical Engineering to log in. This check-in is mandatory.
When the service is completed, the FSE shall document services rendered on a legible ESR(s). The FSE shall be required to log out with Biomedical Engineering and submit the ESR(s) to designated Biomedical Engineering individuals.
CONTRACTOR: Provide the telephone number(s) of your Service Department and contact person:
ADDITIONAL CHARGES: There shall be no additional charge for time spent at the site (during or after the normal hours of coverage) awaiting the arrival of additional FSE and/or delivery of parts.
REPORTING REQUIRED SERVICES BEYOND THE CONTRACT SCOPE: The Contractor shall immediately, but no later than 24 (twenty-four) consecutive hours after discovery, notify the COR (in writing) of the existence or the development of any defects in, or repairs required to, the scheduled equipment which the Contractor considers he/she is not responsible for under the terms of the contract. The Contractor shall furnish the CO and COR with a written estimate of the cost to make necessary repairs.
CONDITION OF EQUIPMENT: The Contractor accepts responsibility for the equipment described in Paragraph 1 in “as is” condition. Failure to inspect the equipment prior to contract award shall not relieve the Contractor from performance of the requirements of this contract.
If new G.E. equipment, software, or components are obtained, either through new purchase, upgrade, or replacement of existing equipment, that equipment may be added (and the replaced equipment deleted) from the service contract as needed upon notification of the Contracting Officer via contract modification. Deleted equipment shall be credited in the full amount if deleted before any maintenance or repair has been performed on it.
TEST EQUIPMENT: Prior to commencement of work on this contract, the Contractor shall provide the VA with a copy of the current calibration certification of all test equipment which is to be used by the Contractor on VA equipment. This certification shall also be provided on a periodic basis when requested by the VA. Test equipment calibration shall be traceable to a national standard.