This opportunity from the U.S. Army Contracting Command - Rock Island (ACC-RI) seeks specialized services for granite table maintenance at the Rock Island Arsenal JMTC. - Government Buyer: - U.S. Army Contracting Command - Rock Island (ACC-RI) - Joint Manufacturing and Technology Center (JMTC) - Services Requested: - Annual inspection, calibration, and lapping/repair of laboratory grade granite surface tables - Compliance with ASME B89.3.7 standards - Affixing calibration stickers and providing Service Acceptance Reports after each calibration - Products/Models: - Multiple granite table models and sizes, including Laboratory Grade AA and A tables - Part numbers: R-68-1439, 19567-M, R-82-791, 22387-M, RIA30627, 19583-M, 19587-M, 19589-M, R-86-1412, 22487-M, R-85-3474, RIA31418, R-86-948, R-59-304, R-85-3475, 23415-M, R-85-3565, P-9299 - Unique Requirements: - Vendor must be ISO 17025 accredited - Adherence to Army installation security protocols and working hours - No specific OEMs or vendors are named in the solicitation - Period of performance is a base year plus four option years (total five years), with annual service cycles

Description

This is a Sources Sought only. 

You MUST send e-mail to the POC (heather.j.petersen.civ@army.mil) that includes the following:  1)  Company Name  2)  Cage Code or EUI number  3)  company POC,  e-mail address,  and phone number  4)  capability statement that outlines you have and can complete the work outlined below.

___________________

Scope of Work (SOW)

1. Purpose: The purpose of this Scope of Work (SOW) is to define a performance-based specification to calibrate granite surface tables for the Rock Island Arsenal's Joint Manufacturing and Technology Center (JMTC).  The work shall be performed by an ISO 17025 accredited vendor shall include the inspection, calibration, and lapping/repair of the tabletops of any that are not within specification.

2. Invoices for services must include a breakout of all costs and must be submitted to the JMTC Representative for review and concurance after services are completed.  This is to verify that the invoices are in line with the contract prices and JTR approved prices.

3. Performance Period: Contractor’s effort will begin immediately after Contract/Task Order award.  This will be for a Base year and Four Option Years for a total of five years.

4. Detailed Specifications: Rock Island Arsenal's JMTC includes the list of granite tables covered by this SOW and is represented below.  

Laboratory Grade AA 36A    R-68-1439    48 X 96         36A    19567-M       60 X 120 36F    R-82-791      60 X 96         36F    22387-M       48 X 72 36F    RIA30627     48 X 96         36S    19583-M       48 X 96         Laboratory Grade A 40A    19587-M       48 X 108 40A    19589-M       48 X 108 40B    R-86-1412    72 X 120 40C    22487-M       48 X 96         40C    R-85-3474    72 X 120 40C    RIA31418     72 X 120 40C    R-86-948      96 X 144 40M    R-59-304      48 X 96 40R    R-85-3475    72 X 120       40W   23415-M       72 X 120 299     R-85-3565    48 X 72 299     P-9299          48 X 72        

4.1  Per ASME B89.3.7 the ISO 17025 accredited contractor shall inspect the granite surface plates and determine both its overall and local flatness.  If necessary, the contractor will lap the working surface back to within specification standards.

4.2 Calibration and any necessary lapping shall be performed on an annual basis.

4.3 After the granite surface plate has been calibrated, the contractor shall affix a calibration sticker to the granite surface plate and provide a Service Acceptance Report (SAR) to the JMTC Representative within three (3) business days of completing the calibration.

5. Working Hours: 0600-1430 hours, Monday-Friday, Government Holidays excluded, unless otherwise authorized, and agreed upon, prior to deviation of the work schedule.

6. Security: The RIA-JMTC is an Army installation subject to Department of Defense safeguards, various precautions, and plant protection measures.  At all times during the execution of this SOW, the contractor will maintain adequate plant protection devices to minimize espionage, sabotage, and other malicious destruction and damage.  The contractor shall comply with all security requirements of the Rock Island Arsenal.

7.   Final acceptance inspection shall be performed by the JMTC Representative to verify that all requirements of this SOW (and any modifications thereto) are satisfied.