Opportunity

SAM #PANHCA26P030639

Sources Sought for Recessed Exam Lighting System and Installation at Fort Carson

Buyer

W40M Medical Materiel Center-West

Posted

April 20, 2026

Respond By

April 24, 2026

Identifier

PANHCA26P030639

NAICS

423450, 335122, 238210

This opportunity seeks sources for a recessed exam lighting system and installation at Fort Carson, Colorado, for the US Army Medical Command. - Government Buyer: - US Army Medical Command - US Army Health Contracting Activity (W40M MRC0 WEST) - Products/Services Requested: - Recessed exam lighting system for clinical use (no OEMs, part numbers, or quantities specified) - Must provide bright, focused, homogenous illumination - Accurate tissue color assessment - Remote aiming capability - Fully recessed ceiling mount - Minimum brightness: 14,000 LUX at 3 meters, 49,000 LUX at 1 meter - Energy efficient - Simultaneous control of multiple light heads - Compatible with standard hospital ceiling heights and electrical systems - Includes wall-mounted and handheld controls - Services: - Installation, inspection, and warranty certification - Direct-to-site setup (inspection for damage, coordination of replacement) - Removal and disposition of existing equipment - Work to be performed in Building 7500, Fort Carson, in several rooms - Unique/Notable Requirements: - Lighting system must not compromise the sterile field for vaginal deliveries - Installation and removal to occur during regular weekday hours (Monday-Friday, 0800-1500) - Excludes electrical wiring, structure, or wall control unit rough-in - Place of Performance: - Fort Carson, Building 7500, Colorado Springs, CO 80913 - Contracting office at JBSA Fort Sam Houston, TX

Description

Clinical Salient Requirements:

The lighting system must be operable without compromising the sterile field required for vaginal deliveries. The light must provide bright, focused illumination deep into the birth canal. The illumination must be homogenous and minimize shadows cast by the clinical team. The light quality must allow for accurate assessment of tissue color.

     Technological Salient Requirements:

Server Requirements: The system must allow for precise, remote aiming of the light beam from a distance via remote or wand.

The light fixture must be fully recessed into a standard ceiling grid. The light must provide a minimum level of brightness at a standard working distance. 14,000 LUX at 3 meters and 49,000 LUX at 1 meter. A single controller must be able to operate multiple light heads simultaneously. The light source must be energy-efficient, long-lasting, and produce minimal heat.

Capture Workstation Requirements: N/A

System Requirement:

<ul>
    <li>It must be compatible with standard hospital ceiling heights and grid systems.</li>
    <li>Must operate on standard hospital electrical systems.</li>
    <li>Must include both a wall-mounted control panel and a handheld wand.</li>
</ul>
</li>

Other Requirements:                                                                           

Installation, inspection and warranty certification of recessed lighting system, single mount or up to two (2) flush ceiling mounted exam system. Includes pre-installation walk through, installation and full functional testing of the system. Does not include electrical wiring, structure or wall control unit rough in. Pricing does not include weekend/after-hours work. Direct-To-Site Setup Service: inspecting products for damage and coordinating replacement if required. Delivery is requested Monday through Friday between the hours of 0800-1500. Installation and removal shall take place during any time of day at address: 1650 Cochrane Circle, BLDG 7500, Fort Carson, CO 80913. Removal of existing equipment from site located in building number: 7500. Room information is below for requested lighting replacement:

RM 3823 RM 3822 RM 3821 RM 3820 RM 3819 RM 3805 RM 3808 RM 3809 RM 3811

Coordination of the disposition of existing equipment is the responsibility of the vendor.

Equipment is shipped fully assembled, or to be assembled on site by vendor. Vendors point of contact to be provided at time of award.