Opportunity
SAM #W911S6-26-Q-A011
U.S. Army Procurement of Teledyne Falcon Series Gas Chromatograph for Dugway Proving Ground
Dugway Proving Ground
March 24, 2026
March 31, 2026
W911S6-26-Q-A011
334516
This procurement opportunity from the U.S. Army at Dugway Proving Ground seeks a specialized Gas Chromatograph (GC) system for safety air monitoring, replacing legacy MINICAMS units. - Government Buyer: - U.S. Army, Mission and Installation Contracting Command (MICC), Dugway Proving Ground - Chemical Test Division, West Desert Test Center - OEM Highlight: - Teledyne Instruments Inc. (manufacturer of the Falcon Series GC Model GS-T) - Products Requested: - 1x Teledyne Falcon Series Gas Chromatograph Model GS-T - Single column, single detector (XSD Halogen Detector) - Quick Change Trap Module (TenaxTA) - Plug & Play Column Module Unit 1 MXT-1701 for Sulfur Detection - Internal sample pump with temperature-controlled vapor sample loop - Integrated NUMAVIEW GC engine with touchscreen display, processor, and storage - Startup kit (fittings, tubing, tooling, pelican case) - Data logging, automatic calibration, visible/audible alarms, heated exhaust, sample bypass - Cycle time of 15 minutes or less - Size and weight constraints for portability - 1x Spare Column Module & Detectors (Part Number: 12903(08-4M)) - Column Assembly (MXT-1701, 0.18mm ID, 0.40 umdf) - Non-Recurring Engineering and Design for Phase III features (integration with Halogen or equivalent Detector) - Packing & Handling, Ground Shipping to Dugway Proving Ground, Utah - Unique/Notable Requirements: - GC must detect halogen-containing chemical agents and compounds without phosphorous or sulfur - Must serve as an alternative to FPD and FID detectors - Delivery to be completed at Dugway Proving Ground, Utah - Fixed-price procurement, no option years - Place of Performance/Delivery: - Dugway Proving Ground, Utah (including West Desert Test Center, BLDG 4223 2nd Street)
Description
A Gas Chromatograph (GC) instrument is needed for air monitoring in replacement of MINICAMS instruments.
Procure one GC instrument with the following specifications:
Detector capable of detecting halogen containing chemical agents and compounds that also do not contain a phosphorous or sulfur element. Detector is an alternate to existing Flame Photometric Detector (FPD) and Flame Ionization Detector (FID) Chromatographic column: MXT-1701 phase Air sampling Quick Change Trap Module with flow rate through the trap of at least 500mL/min External mass flow controller, ability to capture the flow meter readings Oven programmable from ambient to 400℃ Cycle time from injection to injection of 15 minutes or less Visible and audible alarms for system faults and sample concentration detections Integrated GC engine, NUMAVIEW with touchscreen display, integrated processor, and storage. Data logging. Ability to capture instrument set points and measured values such as: sample flow, temperatures, pressures, calibration, concentration, injection type – calibration, service, sample, etc. Data captured locally with the ability to retrieve and use for other reporting or calculating of sample data. Automatic calibration feature. Calibration of the instrument using an averaged single point or multi-point calibration curve. Capable of calibrating and obtaining sample concentrations using the onboard integration engine, and values displayed on the graphical user interface. Heated exhaust to prevent expulsion of water. Detector exhaust must expel water in gas/vapor form elimination water drip from the instrument detector. Sample bypass capability. Ability for continuous sample flow at inlet in a bypass state. A minimum flow of 500mL/min required for bypass capability. GC approximate dimensions - 17” x 11” x 8.5” (width x height x depth) Instrument weight approximately 25 pounds
Shipping: Delivery shall be completed by 30 September 2026 or upon customer notification.