# Quebec SEAO #JD-AI-20260917-2

Procurement of Non-mydriatic Ultra Wide Field Retinal Camera for Santé Québec Laurentides

**Posted:** September 28, 2026
**Respond By:** October 09, 2026
**Identifier:** JD-AI-20260917-2
**NAICS:** 423460, 423490, 339115

Santé Québec Laurentides, a regional health authority in Quebec, is seeking to procure a specialized retinal imaging device for clinical use.
- Government Buyer:
  - Santé Québec Laurentides (regional health authority)
- OEMs and Vendors:
  - Optos Canada (designated as sole provider)
- Products/Services Requested:
  - Non-mydriatic ultra wide field retinal camera
    - Must capture retinal images with pupils as small as 2.0 mm
    - Capable of imaging through grades 1-3 cataracts
    - Single-shot coverage of over 180° of the retina
    - Generates four specific retinal images
    - Supports image montage up to 220°
    - Compatible with EyeVu electronic medical record system (TeckSoft)
    - French software and user interface
    - Includes DICOM licenses, image management platform, virtual server support
    - Adjustable motorized table, no dedicated computer required
    - Annotation and measurement tools
    - Prior installation in other Santé Québec establishments
    - Minimum one-year warranty
    - Ongoing clinical training and robust after-sales support (4-hour callback, 72-hour on-site technician, loaner camera if repairs exceed 72 hours)
    - Health Canada certification, compliance with CAN/CSA-C22.2 NO. 60601-1
  - Option to acquire a second camera with identical specifications
- Unique/Notable Requirements:
  - French-language software and interface
  - Compatibility with EyeVu EMR
  - Demonstrated prior installation in Santé Québec
  - No pricing information to be submitted
  - Suppliers must prove ability to meet all technical and clinical requirements
  - Robust after-sales and clinical training support
- Optos Canada is the only OEM explicitly mentioned and identified as the sole provider capable of meeting all requirements.

### Description

This procurement is for a non-mydriatic ultra wide field retinal camera, with an option to acquire a second similar camera. The camera must meet specific technical requirements including capturing retinal photos with small pupils, imaging through cataracts, covering a wide retinal angle, and compatibility with electronic medical record systems. The procurement is a direct award to Optos Canada, the only supplier meeting the clinical, IT, and technical requirements. Interested suppliers must demonstrate their capacity to meet the requirements without submitting pricing information.

[View original listing](https://seao.gouv.qc.ca/avis-resultat-recherche/consulter?ItemId=296d2752-7920-49e5-a32e-94aa9271057f)